51
With Height Adjustable
Malibu/Sovereign
Bathtub and/or Lifting Arm
Voltage
230 V single phase AC
Frequency
50 Hz
Power consumption
1000 VA
Fuses
External
Internal
230 V 10 A
230 V 6,3 A
Power supply cord
Only replaced by qualified personnel.
Lifting stroke (
Malibu/Sovereign
bathtub)
Malibu 1600 with patient lift
Malibu 1700 / Sovereign with patient lift
300 mm
540 mm (inside) / 500 mm (outside)
585 mm (inside) / 500 mm (outside)
Malibu/Sovereign
bathtub raising time (full load)
33 ± 5 s
Malibu/Sovereign
bathtub lowering time (full load)
Malibu 1600
Malibu 1700
Sovereign
Patient lift for Malibu 1600
Patient lift for Malibu 1700
Patient lift for Sovereign
28 ± 2 s
28 ± 2 s
29 ± 2 s
22 ± 2 s (inside) 20 ± 2 s (outside)
22 ± 2 s (inside) 20 ± 2 s (outside)
25 ± 2 s (inside) 21 ± 2 s (outside)
Cyclic time for lifting arm movement
(up-turn-down, full load)
Malibu 1600 with patient lift
Malibu 1700 with patient lift
Sovereign with patient lift
61 ± 2 s
60 ± 2 s
61 ± 2 s
Maximum duty cycle: Lift actuator chair and
Malibu/Sov-
ereign
bathtub
1 minute ON
9 minutes OFF
Maximum duty cycle: Air Spa
system
15 minutes ON
15 minutes OFF
WARNING
To avoid electric shock, make sure that the equipment is connected to:
• Continuously powered supply mains with protective earth.
• Separate fuse and ground fault circuit interrupter (GFCI)
• A mains disconnection device
All installations must be in accordance with local codes and regulations.
NOTE
All drainage, water and electrical installation work should be carried out by qualified personnel, in accordance