77
Instructions For Use
EN
Appendix 1 - Electromagnetic Compatibility
Make sure the environment in which Team3 is installed is not subject to strong sources of electromagnetic
interference (e.g. radio transmitters, mobile phones).
This equipment generates and uses radio frequency energy. If not installed and used properly, in strict
accordance with the manufacturer’s instructions, it may cause or be subject to interference. Type-tested in a
fully configured system, complies with EN60601-1-2, the standard intended to provide reasonable protection
against such interference. Whether the equipment causes interference may be determined by turning the
equipment off and on. If it does cause or is affected by interference, one or more of the following measures
may correct the interference:
•
Reorienting the equipment
•
Relocating the equipment with respect to the source of interference
•
Moving the equipment away from the device with which it is interfering
•
Plugging the equipment into a different outlet so that the devices are on different branch circuits
WARNING: The use of accessories, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the Team3 as replacement parts for
internal components, may result in increased emissions or decreased immunity of the Team3.
WARNING: The Team3 should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the Team3 should be observed to verify normal operation in the
configuration in which it will be used
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the Team3 including
cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
Guidance and Manufacturer’s declaration - electromagnetic emissions
The Team3 is intended for use in the electromagnetic environment specified below. The customer or the user
of the Team3 should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - guidance
RF emissions
CISPR 11
Group 1
The Team3 uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The Team3 is suitable for use in all establishments, other than
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies