Aquaprobe
®
AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev R
PH 7.00 [31/Jan/17]
PH 4.01 [07/Feb/17]
PH 10.0 [07/Feb/17]
ORP+250 [09/Feb/17]
This tells you that the last successful calibration was January 31
st
for the pH 7.00 point,
February 7
th
for the pH 4.01 & pH 10 points and February 9
th
for ORP. If the date field is
dashed (==/===/==), this means the electrode is either not fitted or has never been
calibrated.
Pushing the left or right arrow keys will toggle this screen with the calibration report screen.
PH7 01.2mV
PH4 Slope:56.8mV/pH
PH10 Slope:56.3mV/pH
ORP Offset:-02.6mV
The calibration report screen displays the calibration report values from the last calibration.
To exit this screen press the
ESC
key.