Calibration
RapidCal
DO 100%
Full Cal
Select Full Cal. This will give you a choice of three electrodes:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
Move the cursor arrow to the electrode you want to reset, then press the
MR
key. If you
select Aux Electrodes, you must press OK first to enter the Aux Electrode selection screen.
Once in that screen, select the Aux electrode you want to reset then press
MR
.
A confirmation screen will be displayed.
Are you sure you
want to restore the
factory calibration
values? [ESC]=NO
If you are sure, press the
OK
key. If you want to change your mind, press the
ESC
key. If
you press OK, you will see a message that says CAL RESTORED.
Once factory calibration defaults have been restored, you
must
carry out a
full calibration
of the electrode in question.
10.6. Calibration Reports
At the conclusion of each successful individual electrode calibration, a single line
Calibration Report is displayed. This report contains the raw output of the electrode under
calibration, uncorrected for temperature.
These values can be recorded and used to track the performance and ageing of the
individual electrodes. Please note however, in order to maximise the value of this feature,
all calibrations must be performed at the same temperature otherwise the recorded values
will not be comparable over time.
10.7. Calibration Data Storage and Retrieval
The Aquaprobe
®
contains its own microprocessor and memory. All calibration data,
including the GLP data, is stored within the Probe’s memory. When a Probe is connected to
a Meter, this data is transferred for display and logging.
This is a major advantage and allows you to use a variety of different Probes with a single
Meter, without the need for re-calibration.
In order to recall the calibration data for a certain electrode on the Meter, first select the
calibration screen for that electrode. If, for instance, you selected pH/REDOX(ORP), the
screen below would be displayed.