Aquaprobe
®
AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev R
6.
Select
pH/REDOX(ORP)
. The screen will change to:
pH 7.00?[01/Jan/17]
PH 4.01?[01/Jan/17]
PH 10.0?[01/Jan/17]
ORP+250?[01/Jan/17]
7.
Move the arrow to the bottom line. If you are using 250mV calibration solution
press the OK key to continue. If you are using 229mV calibration solution, move the
cursor to the right, then use the up/down arrow keys to select +229. When the
correct solution has been selected, press OK. The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up. If the calibration is successful, the counter
will reach 100% and the following screen will be displayed.
Offset: 5.5mV
Calibrating
100%
Press [OK]
The Calibration Report on the top line displays the voltage offset between the ORP
electrode output and the value of the calibration solution at the calibration temperature in
+/-millivolts (mV). During normal operation this offset will be subtracted from the ORP
electrode output to give a corrected ORP display.
This value is stored in the Probe's memory and can be recalled at any time. See section
10.7 Calibration Data Storage and Retrieval
. Press
OK
then
ESC
repeatedly to return to
normal reading mode
.
Remove the Probe from the calibration bottle, rinse thoroughly in fresh water, shake off any
excess and dry the outer sleeve with a soft cloth. Dampen the sponge in the storage cap
with storage solution and fit it to the pH/ORP electrode. ORP calibration is now complete.