Aquaprobe
®
AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev R
2. When the Probe is inserted,
ensure the level of the solution is at lease up to the
minimum insertion line scribed around the Probe sleeve.
If the level is low, the EC
electrodes will not be covered and EC will not calibrate properly. If the level is low,
top up with fresh RapidCal solution.
3. Switch the Aquameter
®
on and wait until all readings are
completely
stable. The
longer you can leave the probe to achieve thermal equilibrium before proceeding,
the better.
4. Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF – 104ºF). The
closer to 25ºC the better.
5. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal
6. Select
RapidCal.
The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until all readings are stable, then it will send the RapidCal command to
the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up. If the calibration is successful, the counter
will reach 100% and the following screen will be displayed.
Calibrating
100%
Press [OK]
When calibration is complete, press
OK
then
ESC
to return to normal reading mode
.
Important
If you have ISE electrodes fitted to your Aquaprobe
®
, you must now remove the rubber
sealing caps from all the ISE electrodes.