APPENDIX C
AUTO SEQUENCE PROGRAMMING
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 88 of 93
44M00
To program this sample sequence, proceed as follows:
1.
Set the operation mode: Press the mode key to CC mode.
2.
Set the range: Press RANGE key to force range 2.
3.
Set LOAD ON/OFF Status: Press Load ON.
4.
Set the current values steps 1-8 and store to memory BANK 3 STATES 1-8.
5.
Press EDIT key of 3300F mainframe.
6.
Press the number 2 key to select F2 as the auto sequence location.
7.
Press up/down key to memory bank 3 and state 1.
8.
Press ENTER key to confirm the sequence memory.
9.
Press up/down key to set the test time for that step (T1+ T2).
10.
Press ENTER key to confirm the sequence step.
11.
Repeat steps 7 to 10 to set auto-sequence steps 1-8.
12.
After setting the final step, press the STORE key.
13.
Press up/down key to 1 to repeat the auto-sequence one time.
14.
Press STORE key to confirm the number of repetitions.
15.
Press TEST key to enter TEST mode.
16.
Press number 2 to select auto-sequence F2.
17.
Press ENTER to confirm selection and start TEST.
18.
The load will now step through the auto-sequence.
The current test waveform can be checked on an oscilloscope as shown below - assuming
that the DC Source can supply the programmed load currents.