CHAPTER 4
REMOTE CONTROL PROGRAMMING
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 70 of 93
44M00
[STATe:] PRESet {SP} {ON |OFF}
Command Syntax:
[STATe:] PRESet {SP} {ON | OFF}{; | NL}
[STATe:] PRESet? {; | NL}
Purpose:
Reads back whether load is in preset mode.
Description:
This command is used to check if the load is in preset mode.
0 = Preset mode OFF, 1 = Preset mode ON
[STATe:] SENSe{SP} {ON | OFF | AUTO}
Command Syntax:
[STATe:] SENSe {SP}{ON | OFF | AUTO }{; | NL}
[STATe:] SENSe? {; | NL}
Purpose:
Reads back whether the sense function is ON or OFF.
Description:
0 = Sense OFF or Sense AUTO
1 = Sense ON
[STATe:] LEVel {SP} {HIGH | LOW} or LEV {SP} {HIGH | LOW}
Command Syntax:
[STATe:] LEVel {SP} {HIGH | LOW }{; | NL}
[STATe:] LEVel? {; | NL}
[STATe:] LEV {SP} {HIGH | LOW}{; | NL}
[STATe:] LEV? {; | NL}
Purpose:
Reads back whether the load is operating at its LOW or HIGH LEVEL.
Description:
In CC, CR, CV or CP operating modes the user can set two LEVELS of load current,
resistance, voltage or power. The load will read back which level it is at:
0 = Load operating at low level
1 = Load operating at high level
[STATe:] DYNamic {SP} {ON | OFF}
Command Syntax:
[STATe:] DYNamic {SP} {ON | OFF}{; | NL}
[STATe:] DYNamic? {; | NL}
Purpose: