CHAPTER 2
INSTALLATION
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 15 of 93
44M00
2.3
Input Fuse
This product is fitted with a mains input fuse. If it needs to be replaced, please adhere to the
following procedure.
CAUTION
BEFORE replacing the fuse you must switch off the unit and mains
power outlet and disconnect the plug of the AC Power cable from the
input socket of the 44M04 chassis.
WARNING
If prior to exchanging the fuse, there is any abnormal noise or odor, do
not use the unit. Please inform your local sales office to organize repair
of the 44M04 chassis.
To avoid the risk of fire or electronic shock the fuse must only be
replaced with same type and rating as the original. Any replacement
fuse used should meet local national safety standards. Any use of an
improper fuse or shorting the Fuse holder is extremely dangerous and
is strictly prohibited.
2.3.1
Fuse Replacement Procedure
1.
Check the rating of the mains input fuse. Replace only with the correct type and rating.
For 100V/115Vac Input use T2A/250V (5*20mm),
For 200V/230Vac Input useT1A/250V (5*20mm)
2.
The AC line fuse is located below the AC line socket (see Fig 2-2). Use a small
screwdriver to remove the fuse holder. Replace the failed fuse with the appropriate
type and rating according to your mains voltage. (See Table 1-2)
3.
Refit the fuse holder and connect the power cord.