AT381x User's Guide
44
5.3.3
Compare
【
Mode
】
The compare mode is only for the primary parameters.
There are three compare modes for the built-in comparators:
Absolute value
∆
Relative value
∆
%
Direct reading value SEQ
Absolute value
∆
and relative value
∆
% mode:
Absolute value
∆
and relative value
∆
% mode is called tolerance mode. Its principle is as
follows:
Figure 5-6
Tolerance mode
Among them
:
Nominal value: The tolerance mode requires input of the nominal value.
●
Includes this point
○
Excludes this point
Absolute value
∆
= measured value – nominal value
Percent
∆
% = (measured value - nominal value) / nominal value × 100%
Figure 5-7
Example of sorting in tolerance mode
●
Includes this point
○
Excludes this point
Sequential mode:
Figure 5-8
Sequential mode: