AT381x User's Guide
100
3128
Primary parameter BIN7 lower
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
312A
Primary parameter BIN7 higher
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
312C
Primary parameter BIN8 lower
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
312E
Primary parameter BIN8 higher
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3130
Primary parameter BIN9 lower
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3132
Primary parameter BIN9 higher
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
4000
Save settings to current file
Fixed value
:
0001
Write-only register, data 2 bytes
4008
Read current file data
Fixed value
:
0001
Write-only register, data 2 bytes
4010
Save settings to the specified file 0000~0009
Write-only register, data 2 bytes
4018
Read specified file data
0000~0009
Write-only register, data 2 bytes
5000
Open circuit full frequency clear
register
Read correction status
Write a fixed value
:
0001
Read
:
0001 in correcting
0000 correction success
FFFF correction fail
Read and write registers, data
occupies 1 register
Once the correction function is
executed, Modbus will disable the
execution of the write command
and only allow the register to be
read.
5008
Short circuit full frequency clear
register
Read correction status
Write a fixed value
:
0001
Read
:
0001 in correcting
0000 correction success
FFFF correction fail
Read and write registers, data
occupies 1 register
Once the correction function is
executed, Modbus will disable the
execution of the write command
and only allow the register to be
read.
5010
Point frequency 1 setting
Frequency value
Floating point number 0
:
Point frequency off
4-byte floating point
number
Read and write registers, data
occupies 2 registers
5012
Point frequency 2 setting
Frequency value
Floating point number 0
:
Point frequency off
4-byte floating point
number
Read and write registers, data
occupies 2 registers
5014
Point frequency 3 setting
Frequency value
Floating point number 0
:
oint frequency off
Read and write registers, data
occupies 2 registers