AT381x User's Guide
32
4.2.2
Auxiliary Bin [AUX] ON/OFF
If you don't need to sort secondary parameters, the auxiliary bin (AUX) can be turned off.
After being turned off, the secondary parameter limit will not be determined during the
measurement.
Procedure for turning on/off auxiliary bin:
Step 1.
Press the [Meas] key
Step 2.
Press the [BIN MEAS] soft key
Step 3.
Use the cursor key to select [AUX] field
Step 4.
Use the soft keys to turn on/off the auxiliary bin
Soft key
Function
OFF
Auxiliary bin if turned off
ON
Auxiliary bin if turned on
4.3
<BIN COUNT> Page
When press [Meas] key and [BIN COUNT] soft key, the <BIN COUNT> page appears.
Figure 4-4
<BIN COUNT> page
The <BIN COUNT> page will display the count result of the comparator.
Settings bar:
Bin count
Information Bar:
The following count values are monitored in the information bar:
Bin1~Bin9 count value
2nd secondary parameter failure number [AUX]
Unqualified number [OUT]
If the condition is set, it needs to be modified in <BIN TABLE> page.
4.3.1
Counter Function [COUNT]
The instrument counts to the bin pass or fail, the maximum count is 999999, the counting
operation stops and the overflow message “------” appears when this value is reached.
Set the Counter:
Step 1.
Press the [Meas] key
Step 2.
Press the [BIN COUNT] soft key to enter <BIN COUNT> page
Step 3.
Use the cursor key to select [COUNT] field
Step 4.
Use the soft keys to set counter function