23
Startup
Warning: Do not apply DC voltage or current to the test terminal, otherwise the instrument
will be damaged.
WARNING: If test a charged device, make sure that its charge is removed before
measuring.
• Test fixtures and cables:
We recommend that users use our test fixtures or test cables, if using test fixtures or cables
made by user or other company may result in incorrect measurements.
Our test fixture contact reeds are silver plated or gold plated. In long-term work (for example,
1~2 years), the plating surface will be worn. It is recommended to replace the new fixtures
when there are obvious deviations in several tests.
Connect test fixture or test cable to four test terminals of Hcur, Hpot, Lcur, and Lpot on the
front panel of the instrument.
Note:
When the test fixture or test cable is not installed, instrument will display unstable
measurement.