101
Modbus
(
RTU
)
Command
4-byte floating point
number
5020
Point frequency open circuit
correction
0001
:
Point frequency 1
0002
:
Point frequency 2
0003
:
Point frequency 3
Read
:
0000 correction success
FFFF correction fail
Read and write registers, data
occupies 1 register
5028
Point frequency short circuit
correction
0001
:
Point frequency 1
0002
:
Point frequency 2
0003
:
Point frequency 3
Read
:
0000 correction success
FFFF correction fail
Read and write registers, data
occupies 1 register
12.2
Fetch Measurement Data
12.2.1
Fetch measurement data [2000H-2003H]
Register 2000~2003 is used for fetching measurement data of the AT381x
Command
:
1
2
3
4
5
6
7
8
01
03
2000
0002
CRC-16
passive
station
read
register
number of
registers
check code
Response
1
2
3
4
5
6
7
8
9
01
03
byte
single precision floating point
number
CRC-16
Obtain the primary parameter measurement result
:
Send:
1
2
3
4
5
6
7
8
01
03
20
00
00
02
CF
CB
passive
station
read
register
number of
registers
check code
Response:
1
2
3
4
5
6
7
8
9
01
03
04
4E
6E
6B
28
A3
E8
01
03
byte
single precision floating point
number
CRC-16
Among them, B4~B6 is measurement data
:
4E6E6B28 stands for 1E9
(
Low position in front
)
Fetch the secondary parameter measurement result
:
Send: