105
Modbus
(
RTU
)
Command
read
register
number of
registers
CRC
Response:
1
2
3
4
5
6
7
01
03
02
00
01
79
84
byte
data
CRC
Data values
:
Data
Function
Description
0000
Manual range
0001
Auto range
0002
Nominal
Select range based on
nominal value
12.3.4
Measurement Speed Register [3003H]
Write
1
2
3
4
5
6
7
8
9
10
11
01
10
30
03
00
01
02
00
01
57
A0
write
register
number of
registers
byte
data
CRC
Response:
1
2
3
4
5
6
7
8
01
10
30
03
00
01
FE
C9
register
number of
registers
CRC
Read
1
2
3
4
5
6
7
8
01
03
30
03
00
01
7B
0A
read
register
number of
registers
CRC
Response:
1
2
3
4
5
6
7
01
03
02
00
01
79
84
byte
data
CRC
Data values
:
Data
Function
Description
0000
Slow speed
0001
Invalid
This value is reserved
0002
Medium speed
0003
Fast speed
12.3.5
Averaging Factor Register [3004H]
Write
1
2
3
4
5
6
7
8
9
10
11