Technical Manual | 808AH V1.2
Anyload Weigh & Measure Inc. @ www.anyload.com | TROUBLESHOOTING
23
5.5
D
IAGNOSTIC
L
IGHTS
LED
Normal
Failure
DISPLAY
If ON, 12V display voltage is present.
If OFF, 12V display voltage has failed.
If OFF check the power supply
If OFF check the fuse on board
CONTROL
If ON, 3V control voltage is present.
If OFF, 3V control voltage has failed.
If OFF check the power supply
USBLINK
USB connection.
For service and manufacturer use only
FUNCTION
Rapid blinking four times a second (1/4sec).
Normal blinking once a second (1 sec).
Slow blinking once every two seconds (2 sec).
Rapid: Automatic mode
Normal: Preset mode
Slow: Manual mode
CLACTIVE
If ON, 20mA loop is in active mode.
If OFF, 20mA loop is in passive mode.
Passive mode is used for the indicators supplying current
Active mode is used for the indicators set in passive mode
RECEIVE
If flashing, data is being received into the buffer.
If OFF, communication has failed.
If OFF check wiring and indicator
Check indicator to be set in continuous mode
WIRELESS
If ON, wireless option is working properly.
It is OFF if no RF module installed.
Not available with this version
5.6
A
DVANCE
D
IAGNOSTIC
Function
Value
Setting
Description
F9.0
Character
XXXXXX
None
Character counter
Shows number of ASCII characters received in buffer
F9.1
Numeric
XXXXXX
None
Numeric counter
Shows number of ASCII numeric received in buffer
F9.2
String
XXXXXX
None
String counter
Shows number of complete strings received in buffer
F9.3
Checksum
XXXXXX
None
Data checksum
Shows memory checksum check failure
F9.4
Signal
XX
None
RF signal strength
Shows wireless signal strength if it is installed
F9.5
Sensor
XXXX
None
Light sensor indicates day light if greater than 600
Shows light sensor value a number between 0 to 1400
F9.6
Version
XX-XX
None
Displays software version
F9.7
Model
XX-XX
None
Displays model number
F9.8
Protocol
XX
None
Reserved
F9.9
Reset
0
1
Abort
Initialize
Alternative setup reset
Resets all setup values to default, except string settings