User Manual fluidlab R-300
Edition 202104
57
Vers.: 1.2
8
Operating and Maintenance Instructions
8.1
Prohibitions on Use
The system should not be used:
•
for medical diagnostics
•
in potentially explosive environments
•
with hazardous substances (corrosive, flammable)
•
with hot liquids (> 60°C)
The area of use is defined by the technical parameters specified in Chapter 9.1.
8.2
Description of Use
8.2.1
User Groups
The
fluidlab
R-300 has been designed to ensure intuitive operation. This measuring
system is intended for professional users, such as laboratory assistants, doctors and
scientists (physicists, biologists, chemists, engineers etc.)
8.2.2
User Expertise
Operating the
fluidlab
R-300 does not require any specialised knowledge. The owner
of the device should designate users who, based on their training, know how to operate
electrical measuring systems. Users should have basic language skills in English, in
order to be able to understand the messages and program texts.
The
fluidlab
R-300 has been designed in such a way that no special product training is
required to operate it.
8.2.3
Service Life
The
fluidlab
R-300 is suitable for 24/7 use. Portable use is limited by the battery life.
The 24/7 operating mode requires the external power supply unit to be connected. If
the battery is low (20% remaining capacity), the battery icon on the user interface menu
bar of the
fluidlab
R-300 will turn red. If you see these indications, you should connect
the device to the power supply unit and charge the battery. If no charging takes place,
the device will shut down automatically when the remaining battery capacity falls to
5%.
The system’s service life is determined to a large extent by th
e battery life. Under ideal
conditions with the system being used as intended, 300-500 charging cycles would be
standard for the built-in battery. Assuming a service life of 5 years, this corresponds to
one charge per week.
Warning!
The device may only be charged and operated using the supplied power supply unit
described in the specifications (Chapter 9.2). If the power supply unit fails, a
replacement power supply unit should be requested from the manufacturer.
Summary of Contents for fluidlab R-300
Page 1: ...User manual fluidlab R 300 ...
Page 3: ...User Manual fluidlab R 300 4 Edition 202104 Vers 1 2 ...
Page 11: ...User Manual fluidlab R 300 12 Edition 202104 Vers 1 2 ...
Page 19: ...User Manual fluidlab R 300 20 Edition 202104 Vers 1 2 ...
Page 53: ...User Manual fluidlab R 300 54 Edition 202104 Vers 1 2 ...
Page 63: ...User Manual fluidlab R 300 64 Edition 202104 Vers 1 2 ...
Page 67: ...User Manual fluidlab R 300 68 Edition 202104 Vers 1 2 ...
Page 69: ...User Manual fluidlab R 300 70 Edition 202104 Vers 1 2 ...