User Manual fluidlab R-300
40
Edition 202104
Vers.: 1.2
Note
:
Saving the calibration curve - If you click on the
in the “Results” screen, you will
be asked whether or not you wish to save the results.
If you tap on “
Save
”, the
calibration curve will be saved under the entered test name in the calibration curve
list
and you will return to the home screen.
If you tap on “
Discard
”, the
measurement will not be saved.
Figure 18:
“Results” screen.
6.2.5.2
Sample Quantification with Created Calibration Curve
1.
Perform an extinction determination of your choice with your samples.
2.
Prepare your samples and the respective blank liquid in the cuvettes and
make the samples ready for spectrometric measurement with the
fluidlab
R-
300
.
3.
Open the slider on the
fluidlab
R-300
,
as described in Chapter 6.2.2.
4.
Insert sample carrier adapter 02 or 03, as described in Chapter 6.2.2.
5.
On the home screen of the
fluidlab
R-300, select the
Calibration Curve
application by tapping on it.
6.
The
“Start a test”
screen is displayed. Select the curve with which you want to
quantify your sample from the created calibration curves in the
“calibration
curve list”
by tapping on it (Figure 19).
Summary of Contents for fluidlab R-300
Page 1: ...User manual fluidlab R 300 ...
Page 3: ...User Manual fluidlab R 300 4 Edition 202104 Vers 1 2 ...
Page 11: ...User Manual fluidlab R 300 12 Edition 202104 Vers 1 2 ...
Page 19: ...User Manual fluidlab R 300 20 Edition 202104 Vers 1 2 ...
Page 53: ...User Manual fluidlab R 300 54 Edition 202104 Vers 1 2 ...
Page 63: ...User Manual fluidlab R 300 64 Edition 202104 Vers 1 2 ...
Page 67: ...User Manual fluidlab R 300 68 Edition 202104 Vers 1 2 ...
Page 69: ...User Manual fluidlab R 300 70 Edition 202104 Vers 1 2 ...