2-1, page 2-0). Press
ENTER
when the data entry is
complete.
Step 14.
Press the BACK soft key to return to the Markers
Menu.
Step 15.
Press the M2 soft key on the Markers Menu to select
Marker 2. This will open the Markers 2nd Level
Menu. Press the ON/OFF soft key to turn M2 on.
Press the EDIT soft key to open the M2 parameter
for data entry. Enter the desired numerical value us-
ing the keypad or the Up/Down Arrow key (Figure
2-1, page 2-0). Press
ENTER
when the data entry is
complete.
Step 16.
Press the MAIN soft key to return to the Main Menu.
Adjusting a Limit
Step 17.
Press the
LIMIT
key (Figure 2-1, page 2-0) to call up
the Scale Menu.
Step 18.
If the Limit needs to be turned OFF, press the LIMIT
soft key on the Scale Menu. If the Limit needs to be
turned ON, press the LIMIT soft key again. This will
open the Limit parameter for entry. If needed, enter
the desired numerical value using the keypad or the
Up/Down Arrow key (Figure 2-1, page 2-0). Press
ENTER
when the data entry is complete.
Step 19.
If you wish to have an audible beep when the trace is
above the limit line, press the OPT soft key from the
Main Menu (page 2-8).
2 - 35
Chapter 2 Operation
Summary of Contents for Site Master S330A
Page 3: ......
Page 15: ...This page is intentionally blank 1 7 Chapter 1 General Information...
Page 23: ...2 7 Chapter 2 Operation Figure 2 3 Site Master Menu Structure...
Page 54: ...2 38 Chapter 2 Operation Figure 2 10 Maximum Distance and Resolution vs Frequency Span...
Page 83: ...3 14 Chapter 3 Screen Capture Program NOTES...
Page 85: ...Color Cover P N 00986 00018...