CAUTION:The selected memory location will be over-
written by the SAVE DISPLAY operation. No warning
is given.
SAVE
SETUP
Saves the current system setup to 1 of 9 internal
non-volatile memory locations. When the key is
pressed, “Save Setup:” appears on the display. Select
an appropriate number from the keypad and press the
ENTER
key to implement.
CAUTION:The selected memory location will be over-
written by the SAVE SETUP operation. No warning is
given.
START
CAL
Opens a calibration menu. Use the Up/Down Arrow
key and
ENTER
key to select the location to store
the calibration (A or B) or cancel the command. The
existing calibration in the location chosen will be
overwritten once the calibration is performed.
Follow the text in the message area that instructs you
to do the following:
q
Connect OPEN, Press ENTER
The Site Master then measures the “open” test
port (standard calibration), or the calibration
“open” (precision calibration kit) that you must at-
tach to the end of the test port or transmission line.
2 - 4
Chapter 2 Operation