Setting Power
Level Step
Size
There are two ways to set the step size of the power
level step that occurs after each frequency
sweep—set the step size or set the number of steps.
The step size range is 0.01 dB (Log) or 0.001 mV
(Linear) to the full power range of the signal genera-
tor; the number of steps range is 1 to 10,000. The
power level step size is set from the Level Sweep
Ramp menu.
Press
Step Size
to open the step size parameter.
Press
Num of Steps
[
PNS
] to open the number of
steps parameter.
Open the parameter you wish to change, then edit
the current value using the cursor control keys, ro-
tary data knob, or enter a new value using the key-
pad and appropriate termination soft-key. When you
have finished setting the open parameter, close it by
pressing its menu soft-key or by making another
menu selection.
3-62
MG369XB OM
Power Level Sweep Operation
Local (Front Panel) Operation
RANGE
This error message is displayed
when (1) the step size value entered
is greater than the level sweep
range or (2) the number of steps en-
tered results in a step size of less
than 0.01 dB (Log) or 0.001 mV
(Linear). Entering a valid step size
will clear the error.