g. Now, return to the Step Sweep menu
by pressing the
< Previous
soft-key.
h. At the Step Sweep menu, press
Dwell Time
to open the dwell-time-
per-step parameter for editing.
i. Set the dwell time to 1 second.
NOTE
Monitor the MG369XB's horizontal
output on the oscilloscope to deter-
mine sweep start and stop.
Step 2.
As the MG369XB steps through the full
frequency range, measure the maximum
and minimum power meter readings and
record the values on the test record. Ver-
ify that the variation (difference between
the maximum and minimum readings)
does not exceed the value noted on the
test record.
Operation Verification
Level Accuracy and Flatness Tests
MG369XB OM
5-13