37XXXC OM
A-29
ALPHABETICAL LISTING
C
MENU
DESCRIPTION
GPIB COMMAND
CONFIRM
CALIBRATION
PARAMETERS
Used for Standard OSL method, coaxial line type.
None
PORT 1 CONN
XXXXXXXX
Calls Menu C4_P1 or C4A_P1, which displays the Port 1
test port connector type to be used during OSL calibration.
This should agree with the connector type that both your
calibration components and the test device mate with.
Move cursor here and press Enter to display menu used to
change connector type.
None
PORT 2 CONN
XXXXXXXX
Calls Menu C4_P2 or C4A_P2, which displays the Port 2
test port connector type to be used during OSL calibration.
This should agree with the connector type that both your
calibration components and the test device mate with.
Move cursor here and press Enter to display menu used to
change connector type.
None
REFLECTION
PAIRING
XXXXXX
Calls menu C13, which lets you select the pairing (mixed or
matched) for the types of reflection devices (open/short)
that you will use on Ports 1 and 2 for calibration.
None
LOAD TYPE
XXXXXXXX
Calls Menu C6, which displays type of load selected for
calibration—broadband fixed or sliding. Move cursor here
and press ENTER to display menu used to change load
type.
None
THROUGH LINE
PARAMETERS
Calls Menu C20, which lets you enter throughline parame-
ters—including offset length and loss equation coefficients.
None
REFERENCE
IMPEDANCE
Calls Menu C17, which lets you choose the reference im-
pedance value (1
mW
to 1 k
W
) for the devices connected to
Ports 1 and 2 for calibration. Default value is 50
W
.
None
TEST SIGNALS
Calls menu CAL_SU2, which lets you enter calibrate Flat
Test Port Power or change source power(s) and attenuator
settings.
None
START CAL
Starts the standard (OSL) calibration sequence using coax-
ial standards.
BEG
PRESS <ENTER>
TO SELECT
OR CHANGE
Pressing the Enter key implements your menu selection.
None
Menu C3, Confirm Calibration Parameters
Summary of Contents for 37 C Series
Page 3: ......
Page 4: ......
Page 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Page 32: ......
Page 41: ......
Page 61: ...Intentionally Blank...
Page 97: ......
Page 103: ......
Page 223: ......
Page 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Page 276: ......
Page 323: ......
Page 567: ......