.Step 4
Choose VIEW DATA from menu (bottom left) and
press
Enter
.
.Step 5
While observing sweep indicator, allow at least two
complete sweeps to occur. (One complete sweep if us-
ing single channel display.)
.Step 6
Choose STORE DATA TO MEMORY from menu
and press
Enter
.
.Step 7
Choose VIEW DATA (/) MEMORY from menu and
press
Enter
.
.Step 8
While observing sweep indicator, allow at least two
complete sweeps to occur. (One complete sweep if us-
ing single channel display.)
.Step 9
Verify that the peak-to-peak High Level Noise falls
within the area between the two limit lines.
NOTE
Displayed data is only valid for the first few sweeps.
.Step 10
Press
Ch 3
key.
.Step 11
Repeat step 4 through 9 for channel 3.
OPERATIONAL CHECKOUT
MILLIMETER WAVE
14-28
37XXXC OM
MENU NO1
TRACE MEMORY
FUNCTIONS
VIEW DATA
VIEW MEMORY
VIEW DATA
AND MEMORY
VIEW DATA (/)
MEMORY
SELECT
TRACE MATH
STORE DATA
TO MEMORY
DISK
OPERATIONS
Summary of Contents for 37 C Series