Operator’s Manual
pg. 10
Rev 5/21/18
4.
Attach the HCl Zinc Filter Assembly (part# H36) to the “Vent Hose” as shown below. Position the hose
and the filter to run downhill from the instrument to prevent condensation from accumulating inside.
Replace the filter every four months or when pH paper placed at the end of
the filter indicates that acid fumes are exiting the filter.
5.
Plug the power cord into a grounded receptacle.
Water Drain
Hose
Vent Hose
HCl Zinc Filter
Assembly (H36)