Disposal
42
Edition 02.19
qTOWER
3
84 / qTOWER
3
84 G
For the setup at the new location observe the notes in section "Installation con-
10.5 Storage
NOTICE
Possible device damage due to environmental conditions and condensation water.
The device must only be stored in air-conditioned rooms. The atmosphere must be low in
dust and free from aggressive vapors.
If the device is not installed immediately after delivery or not required for prolonged pe-
riods, it should be stored in its original packaging. A suitable desiccant should be added
to the equipment to prevent damage from moisture.
The following requirements are placed on the climatic conditions in the storage room of
the qTOWER³ 84:
Temperature range: -10 °C to +55 °C
Max. humidity: 10 % to 30 %
11
Disposal
The operator of the qTOWER³ 84 must dispose the waste materials that occur during
measurements (sample materials) in accordance with the statutory and local regula-
tions.
At the end of its service life, the qTOWER³ 84 and all its electronic components must be
disposed of as electronic waste in accordance with valid regulations.