Raw Data Logging and Exporting
Before starting the measurement by clicking the “Start” button, click the “Record” button to save the raw
data from the AS703x. When a measurement is stopped with the “Stop” button, a pop-up window will
appear to select the file location and save the CSV file.
Figure 52:
Export Raw Data Menu
5.2.6
Signal Optimization
Three settings have a major impact on signal strength and quality:
●
LED current
●
OFE gain
●
Offset compensation
LED current has a direct impact on signal strength with minimal impact on noise. OFE gain will
increase overall signal strength but also increase noise. We recommend the following settings to begin
with and start experimenting from there: