Test the data transmission by using a terminal (AMR-Control,
WIN-Control, DATA-Control, WINDOWS Terminal):
Address the device with its device number
Gxy
(s. man. 6.2.1),
query the programming by
P15
(see man. 6.2.3).
Only check the sending line by cycle input via command
Z123456
and
control in the display.
Test the receiving line by using the key
and monitor control.
Error:
Data transmission within network does not function.
Remedy:
Check that all devices are set to different addresses,
address devices individually via terminal and command
Gxy
,
addressed device is OK when the feedback is at least
y CR LF
. If
data transmission is still not possible, disconnect network. devices,
check devices separately at data cable of the computer (see above).
Check the wiring for a short circuit or twisting.
Are all network distributors supplied with power?
Network and check the devices successively again (see above).
If the device is, after the above inspections, still not performing as specified in
the operating instructions, it must be sent to the factory in Holzkirchen,
Germany, including a short report and possibly control printouts. The software
AMR-Control allows to print the monitor pages including the programming and
also to save the terminal operation and to print it out.
11. ELECTROMAGNETIC COMPATIBILITY
The data acquisition system ALMEMO
®
5590-2 meets the electromagnetic
compatibility (EMC) safety requirements specified in the relevant CE directive
issued by the council for the alignment of legal regulations of the member
states (89/336/EWG).
The following standards have been applied for the evaluation of the product:
EN 50081-1:1992
EN 50082-1:1992
IEC 801-2 8kV, IEC 801-4 1kV
IEC 801-3 3V/m: deviation<100µV
The following notes must be observed when operating the instruments:
1. If the standard sensor cables (1.5m) are extended it must be considered that the
measuring lines are not guided together with power mains and that they are
appropriately shielded to protect against any coupling of disturbance signals.
2. If the instrument is operated within strong electromagnetic fields an additional
measuring error must be expected (<50mV at 3V/m and 1.5m thermocouple
transducers). After the irradiation the device operates again within the specified
technical data.
ALMEMO
®
5590-2
50
Electromagnetic Compatibility