7.3 Cyclic Measuring Point Scan
For cyclic measuring point scans the measuring or output cycle (s. 7.3.1/2)
must be programmed. The measurement is started with the key
START
and the
control lamp
MEAS.CYCLE
will be continuously on. If a peripheral device is
connected the measured values will be provided in cycles and the control lamp
OUTPUT CYCLE
will be on. If the memory is active (see 7.4.1) the measured
values are stored and, in addition, the control lamp
MEMORY
is on.
Start cyclic measuring point scan:
key:
START
The
automatic measuring point scan can be stopped
by operating the key
STOP. The control lamps
MEAS.CYCLE
,
OUTPUT CYCLE
and
MEMORY
turn off.
Stop cyclic measuring point scan:
key:
STOP
7.3.1 Output Cycle
The output cycle, the output channel and the output format (s. 7.2.2) can be set
in the function OUTPUT CYCLE for cyclic measuring point scans and outputs.
Function OUTPUT CYCLE:
OUTPUT CYCLE
Un Outp.cycle 00:30:00
Example:
output channel ´U´ (V24), column format, output cycle 30 min.
The output cycle is programmed using the 6-digit format hh:mm:ss (see 5.4).
Clear output cycle and stop the measuring point scan with using the key
CLEAR.
ALMEMO
®
5590-2
36
Cyclic Measuring Point Scan