Amico Accessories Inc.
Symptom
Possible Cause
Solution
Advanced Interface is blank
• Loose/disconnected cable
• Open the Work Surface (see page 27) and
check connection to the interface
• Cart’s batteries are drained
• Plug in the cart to charge the batteries
• Interface is frozen
• If the cart is equipped with a Reset
Switch, activate switch to reset system.
If the cart is not equipped with a Reset
Switch, briefly disconnect the Interface
Cable from inside of the Main Bin, then
reconnect to force an interface reset.
Automatic Height Adjustment is not
working
• Cart has not yet been initialized
• Initialize the cart (see pages 4, 24-25)
• Batteries have no charge
• Charge the batteries by plugging it into a
nearby outlet
• Cart is already in the lowest/highest
position
• If the cart is already in the highest or
lowest position, it cannot be further
elevated or lowered (respectively).
Reverse the direction of the height
adjustment
• Connection to the interface is loose/
disconnected
• Open the Work Surface (see page 27) and
check the connection to the interface
Both Arrow Keys flashing (STD. Interface)
Home Icon Crossed Out (ADV. Interface)
• Loss of/no home position set
• Home position has been lost
• . Hold the Down Arrow until the cart
reaches bottom (if cart is unable to move
upwards only).
Single Arrow Key (up or down) flashing
(Standard Interface only)
Warning icon appears (Advanced
Interface only)
• Collision detected
• Clear any obstacles preventing height
adjustment. Adjust the cart slightly in
the opposite direction than desired, then
resume use in desired direction.
Difficulty steering the cart
• Casters are sticky/covered in debris
• Casters are locked
• Unlock the casters
• Clean the caster
• Casters are damaged
• Contact Amico Accessories for a
replacement
Difficulty closing the Work Surface
• Devices/objects in the Main Bin are
too large
• Ensure objects in the Main Bin are within
the specified size range
• Lock is engaged
• Ensure the lock is disengaged (turned
counterclockwise) prior to closing the
Work Surface
• FIT Access Door is open
• Ensure the FIT Access Door is closed and
not interfering with the Work Surface
If the above solutions do not solve your symptoms or you are in need of parts/hardware, please contact Amico
Accessories, 1-877-264-2697.
Troubleshooting
Section 8: Troubleshooting, Preventative Maintenance,
Storage, & Cleaning