44 | Model ta5000R Gas Purity Monitor
State Diagram for Super User Mode
This chart illustrates how the menus are arranged in Super User Mode.
Each large circle represents a screen that is displayed on the instrument.
The names printed above each arrow represent a menu label and menu
key. The arrows indicate where pressing the key associated with that label
will take the user.
Shut Down
System shut down for maintenance or for any other reason is very simple.
If shut down is required due to an emergency or for any safety reason,
simply switch the instrument off using the A/C DIN Switch located on the
rear panel of the analyzer (0 depressed is off).
CARRIER GAS TO THE INSTRUMENT MUST NOT BE STOPPED
UNTIL ALL TEM- PERATURES ARE BELOW 50° C.
To shut down the instrument:
From the main menu press the Idle key to stop analysis.
Press the Cool key and allow the instrument to cool down for two to three
hours. Switch the instrument and all other components off.
Shut off the carrier gas.