34 | Model ta5000R Gas Purity Monitor
After pressing Warmup the instrument will perform a series of system
tests and will begin heating critical zones. The results from these tests will
appear on the Warm-up screen. The Warmup menu item will also change
to Cool. If you wish to stop the Warm-up process, press the Cool key and
the instrument will start to cool down. Do not stop the carrier gas until
all the measured temperatures reach below 50
0
C. These temperatures
are displayed on the Temperature and Status screen. If any errors occur
during the Warm-up process the system will stop the Warm-up process.
Proceed to Page 46 Error Messages for more detail.
Warmup Mode Screen
The Warm-up process typically takes 20 to 30 minutes. When the warm-
up process has completed succesfully a
Run
munu label will appear.
Pressing the
Run
key will take the user to the main screen where the Main
menu will be available.
Warmup Pass Screen
WARM-UP
Idle Mode
12:34:25
Module Electronics:
Detector System:
Detector Temperatures:
Carrier Pressure:
Pass
Pass
Pass
Pass
***** Successful Warm-up *****
Run Cool
Page
Warmup Mode
12:35:25
Page
Module Electronics: Pass
Detector System:
Detector Temperatures:
Carrier Pressure:
Cool
WARM-UP
Summary of Contents for TRACE ANALYTICAL ta5000R
Page 14: ...14 Model ta5000R Gas Purity Monitor This page left blank...
Page 23: ...User Manual 23 Display screen...
Page 48: ...48 Model ta5000R Gas Purity Monitor Dimensional Drawings ta5oooR 3 D View...
Page 49: ...User Manual 49 22 115 6 92 ta5oooR Side View...
Page 50: ...50 Model ta5000R Gas Purity Monitor 24 070 16 70 ta5oooR Top View...