User Manual | 43
2. With the asterisk on the top line, one can toggle between single, cycle
and rerun as analysis options. Single will run one event if the “Run” key
is depressed. Cycle will continuously run the loaded event program until
IDLE is depressed. Pressing IDLE in Super User Mode will immediately
stop the cycle. This can result in the valve remain- ing in the inject posi-
tion. If this happens the valve should be manually returned to load posi-
tion (see section 5 below). Rerun will reanalyze the previous chromato-
gram with any database changes and display the new results, although,
the new analysis cannot be printed.
3. Analysis Database (Anal DB) contains the parameters set at the factory
used to integrate peak data. Adjustments to these factory settings are at
the risk of the user.
4. Depressing the “Edit” key next to the Event line will prompt the
Event screen. Program 1 is the factory program automatically loaded
when in the Normal User mode. Any modifications to the event screen
are at the risk of the user.
5. With the asterisk next to the Manual line provides a new menu which
you can toggle the valve and zero the detector for troubleshooting and
service issues. Manual valve actuation should only be performed by an
individual who completely understands the operation of the instrument.
Improper use can result in damage to the instrument in certain applica-
tions.
System Settings
The functionality of this screen is consistent with Normal User.
Summary of Contents for TRACE ANALYTICAL ta5000R
Page 14: ...14 Model ta5000R Gas Purity Monitor This page left blank...
Page 23: ...User Manual 23 Display screen...
Page 48: ...48 Model ta5000R Gas Purity Monitor Dimensional Drawings ta5oooR 3 D View...
Page 49: ...User Manual 49 22 115 6 92 ta5oooR Side View...
Page 50: ...50 Model ta5000R Gas Purity Monitor 24 070 16 70 ta5oooR Top View...