28 | Model ta5000R Gas Purity Monitor
INSTALLATION
INSTRUCTIONS
The installation instructions must be strictly followed for safe and
reliable operation of the analyzer.
If any inconsistencies are encountered during installation, stop im-
mediately. do not continue the installation without direction from
Ametek customer service.
Tools Suggested for Installation
All connections are made using U.S. standard size conectors.
Open end wrenches: 1/4”, 5/16”
Flow Meter (Digital in the sccm range)
Mirror
Magnet
Making Gas Connections
Prepare for Gas Connections
Remove the red plastic shipping cap from the Hg vapor trap (scrubber).
Remove any 1/16” VICI plugs from the rear panel.
Remove all other shipping caps immediately prior to connection of gas
supplies.
Connect Carrier Gas
Matching the carrier gas to the sample will prevent upsets caused by the
elution of the extraneous balance peak. See gas specifications for Carrier
Gas in the Specifications Section for details. The instrument connection is
made at the “CARRIER” 1/16” VICI port on rear panel. Pressurize external
line to 80 psig (Minimum 70 psig, Maxi- mum 90 psig).
Connect Actuator Gas
This analyzer needs high purity inert gas or CDA as valve actuator gas.
See gas specifications for Actuator Gas in Specifications for details.
The instrument connection is made at the “Actuator Gas” 1/8” VICI port
on rear panel. Pressurize external line to 80 psig (Minimum 70 psig, Maxi-
mum 150 psig).
!
CAUTION
!
CAUTION
Summary of Contents for TRACE ANALYTICAL ta5000R
Page 14: ...14 Model ta5000R Gas Purity Monitor This page left blank...
Page 23: ...User Manual 23 Display screen...
Page 48: ...48 Model ta5000R Gas Purity Monitor Dimensional Drawings ta5oooR 3 D View...
Page 49: ...User Manual 49 22 115 6 92 ta5oooR Side View...
Page 50: ...50 Model ta5000R Gas Purity Monitor 24 070 16 70 ta5oooR Top View...