11.
A
spects of calibration and setup
11.1. Calibration setup acc. ANSI
The ANSI standard specify following conditions for verification/calibration:
-
The surge generator is connected to the back filter via the coupling network in the relevant coupling
mode.
-
All the conductors of the mains connection that supply the back filter, including protective ground, are
disconnected from the mains and shorted together at a point upstream from the back filter.