AMETEK CTS
PCD 8 s-series
Operating Manual
V 2.02
23 / 53
5.
PCD 8 s-3-690-1000-100/200
5.1.
Input PCD 8 s-3-690-1000-100/200
Figure 5.2: Input coupling PCD 8 s-3-690-1000-200
1
Decoupling network L1-N
2
Decoupling network L3-L2
3
Sync unit
4
Surge coupling box (SCB)
5
Cooling fans (200 A-models only)
6
Power supply input L1-N / DC+-DC-
7
Power supply input L3-L2
8
Earth rail (PE)
9
Surge generator synch. input
10
Jumper for capacitance setting (Jc)
11
HV-high input from Surge generator
12
Jumper for impedance setting (Ji)
1
Decoupling network L1-N
Surge decoupling network, L1-N
.
2
Decoupling network L3-L2
Surge decoupling network L3-L2
.
3
Sync unit
For synchronisation purpose the synchronisation unit needs to be connected to the decoupling units 1 and 2
(also see 9).
4
Surge coupling box (SCB)
Surge coupling box for IEC coupling L-L and L-PE. An additional SCB is required for ANSI and optional IEC
coupling (in combination with the SCB included in delivery).
5
Cooling fans (200 A-models only)
For operation, the fan motors must be connected to the power supply units supplied, there is no separate on/off
switch for this purpose
.
6
Power supply input L1-N / DC+-DC-
Input screw terminal L1(DC+) and N(DC-). Max. voltage 3 * 690 V AC, 1.000 V DC
.
Summary of Contents for PCD 8 S Series
Page 37: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 37 53 ...
Page 39: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 39 53 9 2 IEC coupling Line to Line 18 µF ...
Page 43: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 43 53 9 6 Optional ANSI coupling 1 5 Basic ...
Page 44: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 44 53 9 7 Optional ANSI coupling 2 5 Basic ...
Page 52: ......
Page 53: ...AMETEK CTS PCD 8 s series Manual of operation V 1 00 53 53 ...