SealBio-2 Semi-Automatic Microplate Heat Sealer Operation Manual Chapter 5 Troubleshooting Guide
─
9
─
Chapter 5 Troubleshooting Guide
Problems and actions
No.
Common problem
Possible cause
Action(s)
1
No display after power on
No power
Check
power
supply
and
plugged properly
On/Off switch broken
Change switch
The fuse is broken
Change fuse
(
5X20 250V 3 A
)
Others
Contact the supplier
2
Display “
Err001
” and alarm
“
Du
…” sound
Over-temperature
Contact the supplier
3
Display “
Err005
”, “
Err006
”,
and alarm “
Du
…” sound
Open or short circuit of
temperature sensor
Contact the supplier
4
Display “
Err601
” and alarm
“
Du
…” sound
Motor locked
Contact the supplier
5
Display “
Err405
” and alarm
“
Du
…” sound
The upper micro-switch
is broken
Contact the supplier
6
Display “
Err412
”
The below
photo electric
switch is broken
Contact the supplier
7
Display “
Err411
”
The upper photo electric
switch is broken
Contact the supplier
8
Display “
Err402
”
The out photo electric
switch is broken
Contact the supplier
9
Display “
Err401
” and alarm
“
Du
…” sound
The enter-photo electric
switch is broken
Contact the supplier
10
“
No
plate
“
displayed,
alarming “Du…”
No microplate or not
place it well
Replace the microplate
11
Instrument cannot run
Controll-board is broken Contact the supplier
Motor is broken
12
Abnormal running noise
Wrong
guide
rail
installed
Contact the supplier
13
Buttons don’t work
Plug connector loose
Contact the supplier
Button broken