background image

CONTENTS

CHAPTER 1

INTRODUCTION

------------------------------------------------------------------------------

1

CHAPTER 2 SPECIFICATIONS

--------------------------------------------------------------------------------

2

1. NORMAL OPERATION CONDITIONS:

----------------------------------------------------------------

2

2. THE BASIC PARAMETERS AND CHARACTERISTICS

------------------------------------------

2

CHAPTER 3 BASIC INSTRUCTIONS

-------------------------------------------------------------------------

3

1. STRUCTURE OVERVIEW

----------------------------------------------------------------------------------

3

2. OPERATION PANEL

----------------------------------------------------------------------------------------

4

3. DISPLAY

--------------------------------------------------------------------------------------------------------

4

CHAPTER 4 OPERATION GUIDE

-----------------------------------------------------------------------------

5

1. SEAL CONDITION SETTING

------------------------------------------------------------------------------

5

2. SEAL OPERATION

-------------------------------------------------------------------------------------------

6

3. HEATING PLATE CLEANING

-----------------------------------------------------------------------------

7

4. OTHER FUNCTION

------------------------------------------------------------------------------------------

8

CHAPTER 5 TROUBLESHOOTING GUIDE

-----------------------------------------------------------------

9

PROBLEMS AND ACTIONS

------------------------------------------------------------------------------------

9

ANNEX:WIRING DIAGRAM

------------------------------------------------------------------------------------

10

DOMINIQUE 

DUTSCHER 

SAS

Summary of Contents for SealBio-2

Page 1: ...Operation Manual Version 2 0 SealBio 2 Semi Automated Plate Sealer HANGZOU ALLSHENG INSTRUMENTS CO LTD D O M I N I Q U E D U T S C H E R S A S...

Page 2: ...he packing list when you first open the instrument packing case If you find there is something wrong with the instrument and the Appendix do contact the vendor or the producer HANGZHOU ALLSHENG INSTRU...

Page 3: ...according to them This product is a normal and an indoor instrument which conforms to Standard GB9706 1 Class I B type Before using the device read the manual carefully This unit is designed for use...

Page 4: ...der improper condition also is not applicable for the damage which caused by repair and refit without authorization The instrument should be put in the place of low temperature little dust no water an...

Page 5: ...ICS 2 CHAPTER 3 BASIC INSTRUCTIONS 3 1 STRUCTURE OVERVIEW 3 2 OPERATION PANEL 4 3 DISPLAY 4 CHAPTER 4 OPERATION GUIDE 5 1 SEAL CONDITION SETTING 5 2 SEAL OPERATION 6 3 HEATING PLATE CLEANING 7 4 OTHER...

Page 6: ...gh 3 Adjustable Sealing Temperature 80 200 4 OLED display screen high light and no visual angle limit 5 Precise temperature timing and pressure for consistent sealing 6 Compact footprint only 7 0 inch...

Page 7: ...ower AC200 240V 50Hz 60Hz 2 The basic parameters and characteristics Display OLED Temperature setting 80 200 Temperature accurate 1 Seal time 0 5s 10s increased by 0 1 seconds Cooling time of heating...

Page 8: ...pter mainly introduces the structure operation keys and displays of the instrument as well as preparatory work before starting Please read this chapter before starting when uses this instrument for th...

Page 9: ...SealBio 2 Semi Automated Plate Sealer Operation Manual Chapter 3 Basic Instructions 4 2 Operation panel 3 Display Drawer Key Seal Key Setting Key D O M I N I Q U E D U T S C H E R S A S...

Page 10: ...nd clockwise rotation to add time Time setting range is 0 5 10 seconds Remark It will confirm the values automatic after 5 seconds 1 3 Setting the pressure function Remark Adjust the pressure accordin...

Page 11: ...actical situation 2 Seal operation After setting up the seal condition seal operation as follows 2 1 Press PLATE In Out key and pull the drawer Then put in the standard plate adapter micro plate and s...

Page 12: ...used by misuse and affect the instrument normal use then should clean the surface of heating plate according to following step 3 2 Press PLATE In Out key make the drawer out switch of the power and pu...

Page 13: ...ure The unit will into Power down mode if there is no action on instrument more than 120 minutes Display screen will turn off power off heater supply and only a weak current supply to the main part of...

Page 14: ...act the supplier 5 Display Err405 and alarm Du sound The upper micro switch is broken Contact the supplier 6 Display Err412 The below photo electric switch is broken Contact the supplier 7 Display Err...

Page 15: ...g Diagram for reference only users won t be informed if there is modification for this diagram Control Board Power Switch Input Power AC 200 240V 50Hz 60Hz Fuse 250V 3A 5X20 Position Sensor Display Bo...

Page 16: ...11 Note D O M I N I Q U E D U T S C H E R S A S...