SealBio-2 Semi Automated Plate Sealer Operation Manual
Chapter 4 Operation Guide
─
6
─
Appendix: Normal seal temperature and time
No.
Item
Seal condition
setting
No.
Item
Seal condition
setting
1
Clear-Seal
170
℃,
2.5S
3
Pierce-Seal
180
℃,
4S
2
Peel-Seal
175
℃,
3S
4
Alfoil-Seal
180
℃,
4S
Above parameter just for reference, Sealing efficiency varies depending on the plate type used.
So users can adjust according to practical situation
2 Seal operation
After setting up the seal condition, seal operation as follows:
2.1 Press “
PLATE In/Out
” key, and pull the drawer, Then put in the standard plate adapter,
micro-plate and sealing film and the weighted sealing flake in turn.
2.2 There is “du…” sound after the temperature reached setting temperature. Press “
RUN
” key,
make the drawer shut up and start to seal. And the drawer will go out of automatic after finished
the seal. Then the unit can do the seal again.
weighted sealing flake
Mircoplate and sealing film
Standard plate adapter
Drawer