SealBio-2 Semi Automated Plate Sealer Operation Manual
Chapter 4 Operation Guide
─
5
─
Chapter 4 Operation Guide
1 Seal condition setting
1.1 Setting the seal temperature
Switch on the instrument. Then press the “
PUSH TO SET
” key for 3 seconds. When the display
of temperature is beginning to glitter, and turn the “
PUSH TO SET
” key to set the seal
temperature. Make counterclockwise rotation to lower temperature and clockwise rotation to
rise temperature. Temperature setting range is 80
℃~
200
℃
.
1.2 Setting the seal time
Press “
PUSH TO SET
” key again after temperature-setting. When display of time is beginning
to glitter, and turn the “
PUSH TO SET
” key to set seal time. Make counterclockwise rotation to
decrease time and clockwise rotation to add time. Time setting range is 0.5~10 seconds.
Remark: It will confirm the values automatic after 5 seconds.
1.3 Setting the pressure function
Remark:
Adjust the pressure according to the seal quality (the trace in the plate after
seal). Turn to high pressure if the trace is light, or turn to low pressure if the trace is
much deep.
Long press
“PUSH TO SET
”, and press “
PLATE In/Out
” keys simultaneity for 3Sec. Enter into
the display of adjust pressure and turn the “
PUSH TO SET
” key to set seal pressure. Make
counterclockwise rotation to decrease pressure and clockwise rotation to add pressure. There
are 1~5 shifts for pressure adjustment. 1 is the lowest pressure and 5 is the highest pressure. 3
is factory default values.
It will confirm the set value automatic if no any action to instruments after 5 seconds.
Temperature
Time
Status display
Counter