SealBio-2 Semi Automated Plate Sealer Operation Manual
Chapter 4 Operation Guide
─
8
─
3.5 Pushing the drawer which aim the slide on the both side into the deepest position slowly
after clean up.
4 Other function
4.1 No plate alarming
When the unit is running, if no plate in the unit, it will display “No plate” and alarming.
4.2 Stand-by mode
The unit will into “stand-by” mode if there is no action on instrument more the 60 minutes and
temperature will down to 60
℃
. By press any button to warm up the unit back to setting
temperature.
The unit will into “Power-down” mode if there is no action on instrument more than 120 minutes.
Display screen will turn off, power off heater supply and only a weak current supply to the main
part of control board. By press any button to warm up the unit back to setting temperature.
4.3 Power-off protection
When there is a sudden power lost during the unit running, the motor will stop. Under this
condition, moving part adjustment is forbidden. It will be automatic reset when power on again
next time.
4.4 Automatic counting function
The counter automatic increase one for sealing work once. Press the “
PUSH TO SET
” key for
10 seconds, the counter can reset (it will be zero).
Heater plate