Camera Link registers
61
Goldeye Camera Link Register Controls Reference V1.0.0
RegSensorCoolingPower
Current TEC cooling power consumption. Negative values indicate that the sensor
is being heated.
RegSensorTemperatureControlMode
Defines the control mode for the thermo-electric cooler (TEC) of the sensor. If set
to
TemperatureControl
, sensor temperature is stabilized to the given setpoint(s)
Register name
RegSensorCoolingPower
Register type
IntReg
Access mode
R
Unit
Milliwatts
Address
0x000B015C
Length (Bytes)
4
Bits used
31..0
Register name
RegSensorTemperatureControlMode
Register type
IntReg
Access mode
R/W
Address
0x000B0004
Length (Bytes)
4
Default
TemperatureControl
Bits used
31..0
Enumeration values:
Value
Name and description
0
Off
No sensor temperature control
1
TemperatureControl
Regulates the sensor temperature in accordance with active values of
other
SensorTemperature
functionalities
2
TemperatureControlTarget
(Only G/CL-008 models)
Regulates the sensor temperature by heating and cooling, aiming to
stabilize it at
SensorTemperatureSetpointValue
of the activated
SensorTemperatureSetpointActive
.
Activates
RegSensorTemperatureTargetSetpoint
that sets the
active
SensorTemperatureSetpointActive