Camera Link registers
101
Goldeye Camera Link Register Controls Reference V1.0.0
RegNUCDatasetActiveTemperature
Shows sensor temperature at acquisition of the data set that is currently indexed
by
RegNUCDatasetActive
. The data set should be selected so that the actual
sensor temperature is close to the reference temperature.
RegNUCDatasetAuto
Controls automatic selection of the
RegNUCDatasetActive
.
The number of distinct reference values is limited by available correction data,
depending on the camera model.
Register name
RegNUCDatasetActiveTemperature
Register type
FloatReg
Access mode
R
Address
0x00090014
Length (Bytes)
4
Bits used
31..0
Units
Degree Celsius (°C)
Register name
RegNUCDatasetAuto
Register type
IntReg
Access mode
R/W
Address
0x00090010
Length (Bytes)
4
Default
Off
Bits used
31..0
Enumeration values:
Value
Name and description
0
Off
The automatic mode is off
1
Once
Auto-NUC occurs until target is achieved, then NUCDatasetAuto
returns to Off
2
Continuous
The non-uniformity correction will continue according to the scene
illumination