7
Press
or
to step through menu options or reposition the cursor.
Press
or
to change the value.
Press
to select the current option.
Press
for 5 seconds to save the displayed value.
SETTINGS OPTIONS (*values that should be set at time of installation)
SETTINGS MENU FLOW DIAGRAM
Text displayed
Description of the menu option
* Set time:
Set the time.
* Set date:
Set the date.
Reset Active Alarm:
Reset an active alarm status and restart the functional test.
View event log:
View a list of historic events, including alarms and recalibration.
The 100 most recent events will be displayed.
Service Mode:
Service Mode ON forces the alarm unit to supply the generator with power for diagnostic
purposes. To end, change to OFF and save. For reasons of safety, Service Mode will automat-
ically exit after one hour.
Set alarm delay:
By default, the delay is set to 5 minutes in order to prevent an alarm in case of
a temporary disruption.
Set inspect freq:
Service and maintenance interval in months.
Perform Recalibration:
This menu option must be used after the generator has been repaired and electronic
components have been replaced. Do not recalibrate at any other time.
EXIT MENU:
Leave the settings menu.
AirMaid
RUNNING
and
401
5 sec.
Set time:
09:00
and
09:00
5 sec.
Set date:
2012 Mar 01
and
2012 Mar 01
5 sec.
Reset
Active Alarm
5 sec.
View event log
Entries: 000
NR: 01
Service Mode
OFF
OFF / ON
5 sec.
Set alarm delay
Delay time: 005m
and
Time: 05m
5 sec.
Set inspect freq
Months: 12
and
Months: 12
5 sec.
Perform
Recalibration
NO / YES
5 sec.
ExIT MENU