7.3 Software installation.
ScanX is designed to be installed by your authorized dealer. The user must provide appropriate and
compliant computer hardware where DBSWIN or an authorized third-party imaging program is installed
to operate ScanX. Smooth operation of the ScanX hardware and software can be affected through
specific hardware and software incompatibility on site at the client despite meeting system requirements.
Where further systems are connected to the PC please note that this may change the system
requirements. Observe the system requirements for all connected systems.
7.5 Electrical connection
Perform the following steps to connect the ScanX to the facility Mains. Before initial start-up, check the supply
voltage with the voltage information on the model identification plate (see also “4 Technical Data”).
1. Make sure that the adapter connector is properly installed. Install the connector if necessary.
Remove the cover on the rear of ScanX.
Connect the 24VDC Power Supply Adapter to ScanX
prior to plugging it into the Mains outlet.
2. Connect the 24VDC Power Supply Adapter Output Connection Cable to the Inlet Power Jack
(item 4) located on ScanX rear panel as shown by the ScanX Rear Panel Connection Ports
illustration.
3. Connect the 24VDC Power Supply Adapter to the Mains outlet.
4. Replace and secure the rear cover making sure that it is correctly fitted.
NOTICE
Safety when making the electrical connection
• The unit may only be connected to a correctly installed socket-outlet
• Do not operate any other systems via the same multi-outlet socket
• Make sure the connection lines to the unit are not subject to any
mechanical tension
INSTALLATION
7.4 Connecting appliances securely
Danger can arise when connecting devices together or to parts of systems (eg through leakage currents)
Only connect appliances together when there can be no danger to the operator or to the patient
Only connect devices when there can be no environmental impairment through such
interconnection
When it is not clear from the appliance data sheets that such connection will cause no
danger, then a qualified expert should be consulted to ensure no danger (eg one of the
product manufacturers)
Only connect peripheral units (eg computer, monitor, printer) which conform to the standard
IEC 60950-1 (EN 60950-1) as a minimum.
Summary of Contents for J1100