45
BASE ASSEMBLY - Solenoid, P/N 43980
INSTRUCTIONS FOR WATER SOLENOID REPLACEMENT
(Phillips screwdriver required)
1. Turn unit upside down and remove Base Cover Panel (6 screws) to access the water
solenoid.
2. Turn unit upside down and remove Base Cover Panel (6 screws) to access the water
solenoid.
NOTE:
Before installing, verify that the solenoid does not contain a flow restrictor (view
through the screen filter). See Detail A
3. Disconnect the power leads (2), ground wire (3) and water inlet tube (4) from solenoid.
Remove existing solenoid and replace with new assembly (6) using machine screws (7),
lock washers (8), and flat washers (9) supplied in kit.
4. Turn base over and re-connect the new water inlet tube (4) to the solenoid connector.
Tighten as required toassure an adequate seal, but do not overtighten or cross thread.
5. Reconnect power leads and ground wire to the new solenoid. Re-assemble unit.
IMPORTANT:
TURN OFF WATER SUPPLY BEFORE
BEGINNING THE REPLACEMENT
PROCEDURE.
SCREEN FILTER
DETAIL A
SOLENOID VALVE WATER INTLET.
CONFIRM THAT THERE IS NOT A BLACK DISK (FLOW
RESTRICTOR) PRESENT BEFORE INSTALLING INTO THE BASE