AT2000 XR Film Quality Troubleshooting
This trouble shooting section is intended for the diagnosis issues relating to film processing
anomalies for the AT2000XR film processors. The itemized list below is a summary of issues, for
the complete diagnosis and/or repair of the issue exhibiting refer to the subsequent pages as
identified.
NOTE:
A major problem contributing to poor film processing quality is the contamination of the processor
chemistry. See AVOIDING CHEMISTRY CONTAMINATION provided on page 37 to prevent
problems.
1.
FILMS ARE TOO DARK or GRAINY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
u
Exposure setting incorrect, chemistry’s not within usage range
2.
FILMS ARE TOO LIGHT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
u
Exposure setting incorrect, chemistry’s not within usage range
3.
FILM DENSITY / CONTRAST IS POOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
u
X-RAY setting, chemistry’s not within range
4.
FILM HAS STREAKING AT ONE END ONLY . . . . . . . . . . . . . . . . . . . . . . . . . . .32
u
Developer rack not square or wet inlet rollers, processor not level
5.
FILM HAS STREAKING CONSISTENTLY REPEATING . . . . . . . . . . . . . . . . . . .33
u
Developer rack not square or needs cleaning, processor not level
6.
FILM HAS STREAKING RANDOMLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
u
Chemistry agitators not operating, inadequate chemistry’s
7.
FILM HAS SPOTS / APPEARS DIRTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
u
Film handling, processor cleanliness, inadequate wash water
8.
FILM HAS “BRANCH-LIKE” ARTIFACTS / SMUDGES . . . . . . . . . . . . . . . . . . . .34
u
Static electricity discharge
9.
FILM HAS SILVER / GREEN / BROWN COLOR . . . . . . . . . . . . . . . . . . . . . . . .35
u
Liquid levels low, fixer agitator not operating, premature processing and
chemistry replenishment
10. FILMS IS WET / TACKY AFTER PROCESSING . . . . . . . . . . . . . . . . . . . . . . . . .35
u
Poor room venting, dryer baffles, chemistry’s not within usage range
11.
FILM EMULSION IS PEELING OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
u
Wash water too hot, chemistry’s not within usage range, wrong film type
12. DEVELOPER TRANSPORT STAINING / DEPOSITS . . . . . . . . . . . . . . . . . . . . .36
u
Contaminated chemistry
13. FIXER TRANSPORT STAINING / DEPOSITS . . . . . . . . . . . . . . . . . . . . . . . . . .36
u
Contaminated chemistry
14. WASH/DRY TRANSPORT STAINING / DEPOSITS . . . . . . . . . . . . . . . . . . . . . .36
u
Dryer baffles missing / misaligned, wash water inadequate