12
OPERATOR CONTROLS
INDICATOR LIGHTS
POWER
Illuminates when the POWER switch (located on top)
is in the ON position.
READY
Illuminates when chemistry has reached SET-TEMP
operating temperature (factory Pre-set at 82° F (28°
C). When READY illuminates, chemistry is at the
proper temperature to process films.
PROCESSING
Illuminates when a film has been inserted and is being
transported through the process cycle.PROCESSING
flashes regularly when the film is feeding past the film
inlet shutter and for 5 seconds after the shutter closes.
When processing extraoral films, wait until PROCESSING stops flashing before inserting the
next film to prevent films from overlaping. When the processor is in a stand - by mode, ready
to accept films for processing, the PROCESSING light will not be illuminated. When the
CLEANING MODE is selected, the PROCESSING light flashes (short - long...)
IMPORTANT:
ENDO is intended to process a
single
periapical size film, and by default, if anoth-
er film is attempted to process after the shutter closes from the 1st film, the
process cycle will change to NORMAL Mode.
KEY PADS
ENDO SPEED
The arrow indicator to the left of the keypad flashes when the A/T 2000XR is in the ENDO
SPEED mode. Film is processed in 2-1/2 minutes.
NORMAL SPEED
The arrow indicator to the left of the keypad illuminates steady when the A/T 2000XR is
in the NORMAL mode. Film is processed in 5-1/2 minutes.
DISPLAY SET - TEMP
The selected chemistry set point temperature flashes on the TEMPERATURE display when
the keypad is pressed and held. Otherwise, the TEMPERATURE display indicates actual
chemistry temperature. If the keypad is pressed and the display does not flash a different
value, the set point and actual chemistry temperatures are the same and READY is illumi-
nated. If this is not the case, see TROUBLE SHOOTING or contact your authorized Air
Techniques Dealer. The SET - TEMP is factory pre-set at 82° F (28° C).
CLEANING MODE
Use this keypad only during cleaning procedures. CLEANING MODE activates an extended
10 minute process cycle. PROCESSING indicator light flashes (short - long...) when in this
mode.
The following provides a brief summary description of the operating controls of the AT2000
XR Film Processor. Refer to the Operator's Manual shipped with the product for detailed
operation informaion for the AT2000 XR Film Processor.