22
Bath Temp
(°
F)
Thermistor
Resistance
60
7658
61
7642
62
7274
63
7089
64
6911
65
6739
66
6569
67
6406
68
6245
69
6092
Bath Temp
(°
F)
Thermistor
Resistance
70
5941
71
5797
72
5655
73
5516
74
5382
75
5251
76
5125
77
5000
78
4882
79
4765
Bath Temp
(°
F)
Thermistor
Resistance
80
4651
81
4540
82
4432
83
4328
84
4225
85
4126
86
4029
87
3936
88
3844
89
3755
Bath Temp
(°
F)
Thermistor
Resistance
90
3668
91
3583
92
3501
93
3420
94
3343
95
3266
96
3193
97
3120
98
3050
99
2982
Notes:
To accurately measure the thermistor’s resistance (Ohms), immerse the probe into a bath of water with a
known temperature. A minimum of 2 minutes immersion time is required to acclimate the probe to the
bath. Measure the resistance at the connector points.
Any other resistance value found at the corresponding temperature which is not within the high and low
limit curves is considered Out of Spec and deemed faulty.
THERMISTOR RESISTANCE CHART