24
important to be sure that the sign
(+/-) shown on the A-D matches the
direction of the Cal mode. If in +Cal
then the sign must be positive. If in
–Cal then the sign must be -.
Rotation of the transducer shaft will
facilitate a change from + to
–
should that be necessary.
9. Percentage of Full Scale is the next
adjustment. Default is 100. Should it
be desired to limit the readable range
of the transducer to be used, use the
RIGHT SELECTION BUTTON
and the
UP or DOWN
arrows to change this
value. Hit
ENT
once the desired value
is displayed.
10. Press
the
DOWN
SELECTION
ARROW
to set the
–Cal
. Confirm
–Cal
by hitting
ENT
. Repeat steps 8 & 9.
11. After mV/V and Full Scale have been
entered, the display returns to the A-D
count screen. Hit
ENT
12. If the value is correct, press
ENT
and
PGM
will show on the screen.
13. Press
the
LEFT
SELECTION
BUTTON
to Save the setting. The unit
will power itself off and will hold the set
value until it is changed in the future or
a different Intellect transducer is
attached to the unit.
Should you have any additional questions, please contact your local
AIMCO-
Auditor™ Series Representative
LIT-MAN861 Rev. 12/2014